SMITHS MEDICAL ASD, INC. CADD-SOLIS AMBULATORY INFUSION PUMP; PUMP, INFUSION, PCA Back to Search Results
Model Number 2110
Device Problem Mechanical Problem (1384)
Patient Problem Insufficient Information (4580)
Event Date 12/01/2023
Event Type  malfunction  
Manufacturer Narrative
One device was returned for analysis.Visual inspection showed a swollen dso seal.The tamper seal was not broken.The event history log (ehl) could not be downloaded due to the error.A functional test was performed and the reported issue was duplicated.The device failed the remote dose test and exhibited error 45639 upon start up.It was determined that the most probable cause was due to the pwa.As a result, the pwa and dso seal were replaced.A service history review identified there was no indication that the complaint was related to a service of the device within the review period.B3 and g4: 510k are unknown.
 
Event Description
It was reported that the pump needed repair.It is unknown if there was patient involvement, no adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-SOLIS AMBULATORY INFUSION PUMP
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key18520280
MDR Text Key333247766
Report Number3012307300-2024-00253
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
UNKNOWN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2110
Device Catalogue Number21-2111-0402-02L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2023
Initial Date FDA Received01/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/23/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-