Catalog Number J2C1711
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/25/2023
Event Type  malfunction  
Event Description
It was reported that the bladder of a small volume infusor ruptured during filling.The device contained saline and 5-fluorouracil (5fu).There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H4: the lot was manufactured between april 11, 2022 to april 13, 2022.H10: one actual device was received for evaluation.A visual inspection was performed, and it was noted that the bladder was ruptured.The ruptured bladder was examined for signs of abnormality and no signs of abnormality were observed on the external and internal surfaces of the bladder.The reported condition was verified.The cause of the reported condition could not be determined; however, possible root cause for bladder rupture may be due to inherent variation in the material from manufacturing, as bladders are manufactured with rubber.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18521988
MDR Text Key333455390
Report Number1416980-2023-07081
Device Sequence Number1
Product Code MEB
UDI-Device Identifier04987456508334
UDI-Public(01)04987456508334
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberJ2C1711
Device Lot Number23D007
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/25/2023
Initial Date FDA Received01/16/2024
Supplement Dates Manufacturer Received01/21/2024
Supplement Dates FDA Received02/15/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-