• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problem Capturing Problem (2891)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Event Description
Related manufacturer reference number: (b)(4), related manufacturer reference number: (b)(4).It was reported, the patient presented with an atrial lead.That oversensed noise, which triggered inappropriate mode switches.And with left ventricular and right ventricular leads, which exhibited high capture thresholds.The atrial lead, left ventricular lead, and right ventricular lead were explanted and replaced.The patient was in stable condition.
 
Manufacturer Narrative
The reported events of inadequate capture and high pacing threshold were not confirmed.A complete lead was returned in two pieces.Visual inspection and analysis of the lead found damages consistent with procedure damage.Electrical testing did not find any indication of conductor fractures or internal shorts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18522880
MDR Text Key332973926
Report Number2017865-2024-00898
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2018
Device Model Number1458Q/86
Device Lot NumberA000005095
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/19/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUADRA ASSURA
Patient Outcome(s) Required Intervention;
Patient Age84 YR
Patient SexMale
-
-