• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH VERTECEM V+ CEMENT KIT; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH VERTECEM V+ CEMENT KIT; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Catalog Number 07.702.016S
Device Problem Fluid/Blood Leak (1250)
Patient Problem Insufficient Information (4580)
Event Date 12/26/2023
Event Type  malfunction  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a synthes employee.G4: device is not distributed in the united states, but is similar to device marketed in the usa.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from japan reports an event as follows: it was reported that patient underwent a posterior spinal fusion (l2) for purulent spondylitis on (b)(6) 2023; l2, 3, s2ai fusion.L2, 3 was used viper prime fs, and only l2 was inserted with the cement in question.The surgeon began filling 12 minutes after cement mixing.Cement was filled in order from left to right in l2, and during cementing to the right, the cement was suspected to be leaking and filling was stopped.While checking the image obliquely, there was a suspicion of leakage into the segmental vein.The surgeon determined that there was no problem because the position of the leaking cement had not changed when he checked the image after some time.Patient was stable.This report is for a vertecem v+ cement kit.This is report 1 of 1 for (b)(4).Additional reports are captured under (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERTECEM V+ CEMENT KIT
Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key18525390
MDR Text Key333157037
Report Number8030965-2024-00942
Device Sequence Number1
Product Code NDN
UDI-Device Identifier07611819376250
UDI-Public(01)07611819376250
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07.702.016S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/26/2023
Initial Date FDA Received01/17/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-