Catalog Number UNKNOWN
Device Problem Backflow (1064)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Event Description
It was reported that unspecified bd infusion set had flow issues the following information was received by the initial reporter with the verbatim.Kindly review the below customer concern: ¿rn on icu shared that in room 2215 the day before today (b)(6), she received an ail alarm and when addressing alarm, noted the line to have alternating sections of air (about two inches) and fluids throughout.She stopped infusion and disconnected, she was able to withdraw blood from the patient¿s central line and did not note any air.Rn states this did not affect the patient and no patient involvement was noted.Rn noticed the tubing above pump module was twisted with a secondary line that had an empty bag.Rn stated she feels it was pulling fluids from the primary and air from the secondary.Rn states she did not get an ail alarm until the air was already down the length of the tubing.¿.
 
Manufacturer Narrative
B3.The date received by manufacturer has been used for this field.H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNSPECIFIED BD INFUSION SET
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18525887
MDR Text Key333033918
Report Number9616066-2024-00015
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-