Titan touch pump, cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on the shorter exhaust tube and the inlet tube of the pump.No functional abnormalities were noted with the pump.No functional abnormalities were noted with the either cylinder 1 or cylinder 2.No functional abnormalities were noted with the reservoir.The information received indicated the device had a cylinder-pipe joint failure, but because no functional abnormalities were noted with the returned components, the complaint could not be confirmed as reported.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|