SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number GF-UCT180
Device Problem Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported to olympus, there was too much pressure in the air water channel of the evis exera ii ultrasound gastrovideoscope during automated endoscope reprocessing treatment.The device was returned for evaluation.During the device evaluation, whitish material was found clogging the nozzle.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the evaluation found too much pressure in the air-water channel during aer treatment, bending section cover glue is separated, scratching marks on the key top 1 rubber and the acoustic lens surface, and the nut on the elevator channel plug is damaged.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the observed foreign material in the nozzle appeared to be a solid, whitish material, similar to cleaning agent residue but otherwise could not be identified.A definitive root cause of the issue could not be determined, as there was no deformation observed that might result in the retention of foreign material, however, device reprocessing could not be completed due to the clogged nozzle.The issue was likely the result of user handling/mishandling.The issue may be detected/prevented by following the instructions for use sections below: chapter 7cleaning, disinfection, and sterilization procedures chapter 8 reprocessing workflow for endoscopes and accessories chapter 9 reprocessing the endoscope(and related reprocessing accessories) olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18529322
MDR Text Key333226929
Report Number3002808148-2024-00521
Device Sequence Number1
Product Code ODG
UDI-Device Identifier04953170356346
UDI-Public04953170356346
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/29/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/12/2024
Initial Date FDA Received01/17/2024
Supplement Dates Manufacturer Received01/30/2024
Supplement Dates FDA Received02/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/25/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-