US SURGICAL PUERTO RICO PREMIUM SURGICLIP III; CLIP, IMPLANTABLE Back to Search Results
Model Number 133650
Device Problem Mechanics Altered (2984)
Patient Problem Insufficient Information (4580)
Event Date 11/20/2023
Event Type  malfunction  
Event Description
According to the reporter, during a plate removal procedure, issues were experienced with the loading or firing of the clips.A new clip applier was used to resolve the issue.
 
Manufacturer Narrative
D10 concomitant product: 133650 - 133650 premium surgiclip iii 9.0, lot# p3d0238 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the instrument was cycled, and the pusher was not advancing and did not load the clip.The instrument was disassembled, and the ratchet assembly was damaged, and pusher bar was disengaged indicating the handles were forced open prior to completing the firing cycle.It was reported that the clips not loading properly.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the user attempts to either pull apart the handles prior to completing the handle compression or attempting to squeeze the handles prior to fully releasing the handles after completing a handle compression.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the device contains a ratchet mechanism to ensure that a clip is not allowed to release until a full handle squeeze is applied.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PREMIUM SURGICLIP III
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18531105
MDR Text Key333349369
Report Number2647580-2024-00369
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10884521517394
UDI-Public10884521517394
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K142869
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number133650
Device Catalogue Number133650
Device Lot NumberP3C0633
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2023
Initial Date FDA Received01/17/2024
Supplement Dates Manufacturer Received02/22/2024
Supplement Dates FDA Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SEE H10
-
-