Model Number 102096-81A |
Device Problem
Material Separation (1562)
|
Patient Problem
Discomfort (2330)
|
Event Date 12/19/2023 |
Event Type
Injury
|
Event Description
|
On december 19, 2023, senseonics was made aware of an adverse event where a sensor broke during the removal procedure.
|
|
Manufacturer Narrative
|
When the hcp first pulled the sensor out with the clamp, only the bottom part with the ring was removed.When the hcp went back in to remove the other piece, some dark tissue was removed before removing the rest of the sensor.The sensor was removed in its entirety.The manufacturer is currently performing an investigation and will provide the results with the supplemental report.
|
|
Manufacturer Narrative
|
Sensor was received, a visual inspection showed that the end cap was separated from the sensor.All sensor pieces were successfully removed from the patient.The root cause for sensor breakage is usually excessive force on the removal clamps or grabbing an extremity of the sensor.B4.Date of this report 01 march 2024.G3.Date received by the manufacturer? 01 feb 2024.H3.Device evaluated by manufacturer?yes.
|
|