The most probable root causes associated with this failure mode are disconnected, faulty or damaged components or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The device history records (dhrs) for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
A bent sensor tip was reported with the abbott diabetes care (adc) device.Customer was unable to apply the device and monitor glucose levels.As a result, customer experienced a loss of consciousness and was unable to self-treat, requiring third-party treatment of "insulin" (type/dose unspecified) by a non-healthcare professional.There was no report of death or permanent injury associated with this event.
|
(b)(6) has been returned and investigated.Visual inspection has been performed on the sensor and no issues were observed.The watermark was observed at the base of the tail indicating the sensor being properly inserted.Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.No malfunction or product deficiency was identified.Therefore, issue is not confirmed.An extended investigation has been performed for the updated serial number and the reported complaint and it has been determined that there was no indication that the product did not meet specification.The dhrs (device history review) for the libre sensor and libre sensor kits were reviewed and the dhrs showed the libre sensor and libre sensor kits passed all tests prior to release.Section d4: (serial number) was updated from (b)(6) to (b)(6) based on returned product download.All pertinent information available to abbott diabetes care has been submitted.
|
A bent sensor tip was reported with the abbott diabetes care (adc) device.Customer was unable to apply the device and monitor glucose levels.As a result, customer experienced a loss of consciousness and was unable to self-treat, requiring third-party treatment of "insulin" (type/dose unspecified) by a non-healthcare professional.There was no report of death or permanent injury associated with this event.
|