Model Number 439888 |
Device Problems
Collapse (1099); Fracture (1260); Capturing Problem (2891); Impedance Problem (2950); High Capture Threshold (3266)
|
Patient Problems
Bradycardia (1751); Dizziness (2194); Syncope/Fainting (4411); Asystole (4442)
|
Event Date 12/20/2023 |
Event Type
Injury
|
Event Description
|
It was reported that a lead integrity alert (lia) triggered on the right ventricular (rv) lead two days in a row due to high rate non-sustained episodes and short ventricular intervals.A lead noise warning had triggered in between the two lias.The lead was noted to be oversensing which led to pacing inhibition and detection of ventricular fibrillation (vf) which led to inappropriate shock.The lead was also reported to exhibit high undefined pacing impedance, a pacing impedance spike, high and rising pacing thresholds, and myopotential oversensing.It was reported that the pacing portion of the lead was confirmed to have fractured.The rv lead was explanted and replaced.It was also reported that the right atrial (ra) lead was oversensing and inhibiting pacing.The ra lead was explanted and replaced.It was also reported that the left ventricular (lv) lead exhibited rising impedance, high and rising thresholds, and had fractured.During laser extraction, the lead was noted to have broken due to clavicle crush.The lv lead was partially ex planted and replaced.The patient was noted to have experienced asystole, bradycardia, dizziness and syncope.An entire new system was then implanted.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|