• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SENTINEL CEREBRAL PROTECTION SYSTEM (US); EMBOLIC PROTECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SENTINEL CEREBRAL PROTECTION SYSTEM (US); EMBOLIC PROTECTION DEVICE Back to Search Results
Model Number CMS15-10C-US
Device Problems Break (1069); Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/03/2024
Event Type  malfunction  
Event Description
It was reported that filter break occurred.A transcatheter aortic valve replacement procedure was being performed and sentinel cerebral protection system (cps) was selected for use.The patient's anatomy contained moderate to severe calcified arteries and a tortuous right subclavian artery.The physician decided to pre-wire into the ascending aorta using a long, non-boston scientific (bsc) 0.014mm guidewire.The sentinel cps was advanced over the non-bsc guidewire.Difficulty was encountered in obtaining proper position and the proximal filter of the sentinel cps was deployed three or four times.The proximal filter then failed to deploy, so the sentinel cps was removed from the patient's anatomy.Once outside of the patient, an attempt was made to deploy the proximal filter, which deployed successfully.However, upon inspection, the proximal filter wire loop had partially disconnected from the proximal filter.A new sentinel cps was successfully used to complete the procedure.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTINEL CEREBRAL PROTECTION SYSTEM (US)
Type of Device
EMBOLIC PROTECTION DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18535670
MDR Text Key333205905
Report Number2124215-2024-00488
Device Sequence Number1
Product Code PUM
UDI-Device Identifier00863229000004
UDI-Public00863229000004
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCMS15-10C-US
Device Catalogue NumberCMS15-10C-US
Device Lot Number0032474395
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/03/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/21/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-