• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH A

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. HEMODIALYSIS BLOODLINES; SET, TUBING, BLOOD, WITH A Back to Search Results
Catalog Number SL-2000M2095L
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2023
Event Type  malfunction  
Event Description
As reported by the user facility: event 1.Brief inquiry description: venous pod dysfunction - breakthrough.Detailed inquiry description customer reports at least five incidents of blood and/or saline breakthrough noted by clinical team to machine from venous pressure monitoring line.Upon investigation, customer noted that there is not a diaphragm in one of the venous pods.No injury reported.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).Event 1.The investigation is ongoing at this time.A follow-up will be submitted when the investigation results become available.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMODIALYSIS BLOODLINES
Type of Device
SET, TUBING, BLOOD, WITH A
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
824 12th avenue
bethlehem 18018
Manufacturer (Section G)
B. BRAUN MEDICAL INC.
824 12th avenue
bethlehem PA 18018
Manufacturer Contact
jonathan severino
901 marcon blvd
allentown, PA 18109
4847197287
MDR Report Key18536066
MDR Text Key333150163
Report Number2521402-2024-00009
Device Sequence Number1
Product Code FJK
UDI-Device Identifier04046955348909
UDI-Public(01)04046955348909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K080807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSL-2000M2095L
Device Lot NumberA2200647
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-