Model Number 0998-00-0800-55
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/09/2024
Event Type  malfunction  
Event Description
It was reported that while in stand by mode, the cardiosave intra-aortic balloon pump (iabp) helium tank empties.There was no patient involved.
 
Manufacturer Narrative
Updated fields: (event site postal code - (b)(6).A getinge field service engineer (fse) evaluated and stated that high helium leak coming from the high pressure reducer.Fse replaced helium high pressure regulator.After replacement, leaks fell within the standard parameters.Operational tests carried out with positive results.The unit was cleared for clinical use.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID W/ E/F PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18536577
MDR Text Key333348540
Report Number2249723-2024-00228
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108414
UDI-Public10607567108414
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-55
Device Catalogue Number0998-00-0800-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/09/2024
Initial Date FDA Received01/18/2024
Supplement Dates Manufacturer Received05/23/2024
Supplement Dates FDA Received05/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/10/2019
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-