• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA CRT-D QUAD RF HV; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ASSURA CRT-D QUAD RF HV; NO MATCH Back to Search Results
Model Number CD3365-40Q
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2024
Event Type  Injury  
Event Description
It was reported that patient presented in clinic for routine follow-up.Upon interrogation, it was found that patient's implantable cardioverter defibrillator (icd) was in backup mode due to event queue overflow reset.The icd was able to reset to normal setting by programming intervention.There were no patient consequences.
 
Event Description
New information received notes that the device was explanted and replaced.
 
Manufacturer Narrative
The reported event of backup operation was confirmed.The device went into backup mode due to an event queue overflow.The device was restored by reloading product code in the field.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.The cause of the event queue overflow was determined to be the integrated circuit (ic) in the radio frequency (rf) module.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA CRT-D QUAD RF HV
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18536628
MDR Text Key333211334
Report Number2017865-2024-01434
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508292
UDI-Public05414734508292
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2017
Device Model NumberCD3365-40Q
Device Lot NumberA000008580
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/02/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ATRIAL LEAD; QUARTET; RIGHT VENTRICULAR LEAD
Patient Outcome(s) Required Intervention;
Patient Age89 YR
Patient SexFemale
-
-