Model Number DSX500T11C
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Manufacturer Narrative
H3 other text : device evaluated by third party service center.
 
Event Description
A dreamstation auto cpap device was returned to the third-party service center as part of the recall process with no initial complaint.There was no report of patient harm or injury.The device was evaluated by the service center.The internal part of the device was inspected visually, and corrosion found inside the device.The device has been found unusable and was scrapped.
 
Manufacturer Narrative
The manufacturer received information in relation to a dream station auto cpap unit.The device was returned to a third-party service center.During visual inspection of the device, it was determined the power connector of the unit was corroded.In addition, there was contamination from connector oxide.Device was scrapped at customer's request.Analysis of past events has not indicated an adverse event has occurred due to corrosion.Review of the risk file indicates the potential for a serious adverse event occurring as a result of this incident is unlikely.Additionally, the risk file indicates that corrosion will not substantially affect the performance of the device.This complaint is considered not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINIOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18537622
MDR Text Key333165134
Report Number2518422-2024-03326
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500T11C
Device Catalogue NumberDSX500T11C
Was Device Available for Evaluation? No
Date Manufacturer Received12/20/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/10/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No