MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number FRX500S14
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/22/2023
Event Type  malfunction  
Event Description
The manufacturer was contacted in reference to dreamstation auto device.There was no report of patient harm or injury.The device was returned to third party service center.During the evaluation, the device was visually inspected and found power supply corroded.The device was scrapped.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.
 
Manufacturer Narrative
H3 other text : device not returned to manufacturer.
 
Manufacturer Narrative
The manufacturer received information in relation to a dreamstation auto unit.The device was returned to a third party service center.During visual inspection of the device, it was determined the power connector of the unit was corroded.Device was scrapped at customer's request.Analysis of past events has not indicated an adverse event has occurred due to corrosion.Review of the risk file indicates the potential for a serious adverse event occurring as a result of this incident is unlikely.Additionally, the risk file indicates that corrosion will not substantially affect the performance of the device.This complaint is considered not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18537632
MDR Text Key333165541
Report Number2518422-2024-03364
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberFRX500S14
Device Catalogue NumberFRX500S14
Was Device Available for Evaluation? No
Date Manufacturer Received12/22/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-