SMITHS HEALTHCARE MANUFACTURING S.A. DE C.V. MEDEX TRANSTAR PRESSURE MONITORING TRANSDUCER, 84 IN; PROBE, BLOOD-FLOW, EXTRAVASCULAR
|
Back to Search Results |
|
Catalog Number MX9505T |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2023 |
Event Type
malfunction
|
Event Description
|
It was reported the device had no data displayed.It was replaced with a new set, and the data can be read normally.There was patient involvement, but no harm/adverse event was reported.
|
|
Manufacturer Narrative
|
H3: device not received by manufacturer.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
D9: 29-jan-2024.Device evaluation: two used devices were returned for investigation.One (1) out of two (2) returned samples was found rejected during electrical testing.Vacuum, air leak and water leak testing was passed by both devices.Visual inspection of the electrical board found the sample was not conforming due to a lack of solder on the electrical board.The complaint was confirmed.A review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.Awareness for failure ¿while the product was being used, there was no data being displayed¿ was given to personnel who perform the assembly process.
|
|
Event Description
|
Two (2) sets with the same lot and item numbers were reported faulty.
|
|