Upon receipt, the lead under complaint was subjected to an extensive analysis.The performance of the lead was scrutinized, including a visual, mechanical and electrical inspection.The analysis showed that the wires of the conductor coil leading to the lead tip and all ring electrodes were found fractured approximately 42.5 cm distal to the is4 connector pin.The damage is assumed to be the root cause of the clinical observations.Based on the characteristics of this damage, it is reasonable to assume that the lead had been subject to severe mechanical stress in the implanted state.Furthermore, the conductor coil and insulation were found squeezed and deformed approximately 10, 19, 24, 27 and 38 cm distal to the is4 connector pin.These deformations probably resulted from the explantation procedure.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
|