MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 367989
Device Problem Material Protrusion/Extrusion (2979)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2023
Event Type  malfunction  
Event Description
It was reported that during use with the bd vacutainer® sst¿ blood collection tube, there was a bur on the bottom of the tube.There was no report of patient impact.
 
Manufacturer Narrative
E1.(b)(6) hospital.H3.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Manufacturer Narrative
The following fields have been updated with corrected and/or additional information.D9: device available for evaluation: yes.D9: returned to manufacturer on: 16-jan-2024.H.6.Investigation summary: bd received 1 returned sample and 3 photos from the customer for investigation for molding defect.Visual examination of the samples and photos revealed flash on the tubes.Device history record review of reported incident lot determined all product release requirements were met.There were no related quality issues during product manufacture.This complaint has been confirmed for flash.Bd was unable to identify a root cause for indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that during use with the bd vacutainer® sst¿ blood collection tube, there was a bur on the bottom of the tube.There was no report of patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18540222
MDR Text Key333493224
Report Number1024879-2024-00015
Device Sequence Number1
Product Code JKA
UDI-Device Identifier50382903679893
UDI-Public(01)50382903679893
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
BK050036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date04/30/2024
Device Catalogue Number367989
Device Lot Number3116893
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-