• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WALDEMAR LINK GMBH & CO. KG LINKSYMPHOKNEE; ARTICULATING SURFACE, SYMMETRIC, SZ. 3-4, UHMWPE PS, FIXED, H= 12 MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WALDEMAR LINK GMBH & CO. KG LINKSYMPHOKNEE; ARTICULATING SURFACE, SYMMETRIC, SZ. 3-4, UHMWPE PS, FIXED, H= 12 MM Back to Search Results
Model Number 880-213/12
Device Problems Difficult or Delayed Positioning (1157); Positioning Failure (1158); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Failure of Implant (1924); Rupture (2208); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2023
Event Type  Injury  
Event Description
During the surgery, after implanting the tibial component, the surgeon tried to insert the poly component which struggled to enter, and above all, did not fix with the classic snap.Thinking that the pe was faulty, he opened a second identical one, but the problem repeated itself.Not knowing what to do, nobody of us was in the operating room, a third component of a higher thickness as opened but again he had difficulty inserting the component and didn't hear the usual clacking.At this point, hoping that the insert was fixed and stable, but with some concern, the surgery was finished.[customer].
 
Manufacturer Narrative
The review of the device history record showed no deviations.The product complies with the specifications valid at the time of manufacture.Note on exceeding the reporting deadline.Reporting deadline 29 december 2023.Reporting date 19 january 2024.This initial report is delayed due to an it security-related incident from (b)(6) 2023.Due to the incident, we did not have access to our it systems and were temporarily unable to meet our obligation to report incidents on time.The process-relevant functions have been available again since 16 january 2024.Accordingly the reports will now be submitted retrospectively.
 
Manufacturer Narrative
This is the final supplemental report.The complaint is closed.
 
Event Description
During the surgery, after implanting the tibial component, the surgeon tried to insert the poly component which struggled to enter, and above all, did not fix with the classic snap.Thinking that the pe was faulty, the surgeon opened a second identical one, but the problem repeated itself.Not knowing what to do, none of us was in the or, a third component of a higher thickness was opened but again, the surgeon had difficulty inserting the component and didn't hear the usual clacking.At this point, hoping that the insert was fixed and stable, but with some concern, the surgery was finished.[customer].
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINKSYMPHOKNEE
Type of Device
ARTICULATING SURFACE, SYMMETRIC, SZ. 3-4, UHMWPE PS, FIXED, H= 12 MM
Manufacturer (Section D)
WALDEMAR LINK GMBH & CO. KG
barkhausenweg 10
hamburg, hamburg 22339
GM  22339
Manufacturer Contact
lisa effe
barkhausenweg 10
hamburg, hamburg 22339
GM   22339
MDR Report Key18542324
MDR Text Key333228675
Report Number3004371426-2024-00002
Device Sequence Number1
Product Code HRY
UDI-Device Identifier04026575252961
UDI-Public04026575252961
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K202924
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Model Number880-213/12
Device Catalogue Number880-213/12
Device Lot Number2015096
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/29/2023
Initial Date FDA Received01/19/2024
Supplement Dates Manufacturer Received11/29/2023
Supplement Dates FDA Received05/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-