• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE G PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-52
Device Problem No Display/Image (1183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2024
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that, the cardiosave intra-aortic balloon pump (iabp) screen goes off.There was no patient involvement.
 
Manufacturer Narrative
Updated fields: b4, d9, g3, g6, h2, h3, h4, h6 (type of investigation, investigation findings and investigation conclusions), h10, h11 corrected fields: d5, e2, e3, g2 additional information: event site postal code: (b)(6).A getinge field service engineer (fse) on (b)(6) 2024 went onsite and found that the unit could not enter the service mode.He will provide a temporary quotation for the acknowledgement so as to proceed with the repair in advance.(b)(6) 2024 troubleshoot the problem and found that the screen issue is resolved.However, the pcba video generator will be put on hold and unit for monitoring.Temperature sensor replaced and performed all manifold and compressor output tests.Unit ok.Will perform overheat test on (b)(6) 2024 as agreed with bme for 3 days before closing this case.(b)(6) 2024 replaced the screen with a loaner from getinge.The machine was run from (b)(6) 2024 to (b)(6) 2024 and no issue was found.The service mode was also randomly tested at different intervals for 10 times and no issue found.Final quotation amended and issued.Will recommend to replace those parts stated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE G PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18542468
MDR Text Key333231375
Report Number2249723-2024-00225
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108438
UDI-Public10607567108438
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-52
Device Catalogue Number0998-00-0800-52
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/10/2024
Initial Date FDA Received01/19/2024
Supplement Dates Manufacturer Received05/21/2024
Supplement Dates FDA Received05/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2014
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-