• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD POLYSORB; ENDOSCOPIC TISSUE APPROXIMATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD POLYSORB; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number 170052
Device Problems Break (1069); Human-Device Interface Problem (2949)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2023
Event Type  malfunction  
Event Description
According to the reporter, during a laparoscopic rectal prolapse repair procedure, there was difficulty in getting the needle to release.The needle broke as it got out of the handle.A new device was used to resolve the issue.There was no patient injury.
 
Manufacturer Narrative
D10 concomitant product: 173016 endo stitch instrument (lot#: j3h3485ey).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there was difficulty in getting the needle to release.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when the needle is not loaded properly or when the needle is forced into an obstacle.This condition may also occur if excess pressure is exerted on the needle or the attached suture while the jaws are in the open position.In these situations, the needle may flex and ultimately break.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: ensure that the toggle levers are fully retracted prior to opening the jaws of the device.The reloading button should never be pressed when the instrument is in the body cavity, as this will release the needle.Inspect the application site to ensure hemostasis.Place additional sutures or use electrocautery if necessary to complete hemostasis.Endoscopic procedures should be performed only by physicians having adequate training and familiarity with endoscopic techniques.A thorough understanding of the operating principles, risks versus benefits, and the hazards involved in utilizing an endoscopic approach is necessary to avoid possible injury to the user and/or patient.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Type of Device
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DR   0101
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18546007
MDR Text Key333276746
Report Number9612501-2024-00194
Device Sequence Number1
Product Code OCW
UDI-Device Identifier20884521126746
UDI-Public20884521126746
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K934738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number170052
Device Catalogue Number170052
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/02/2024
Initial Date FDA Received01/19/2024
Supplement Dates Manufacturer Received02/08/2024
Supplement Dates FDA Received03/05/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age91 YR
Patient SexFemale
Patient Weight58 KG
-
-