Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2024
Event Type  malfunction  
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) , that the cardiosave intra-aortic balloon pump (iabp) battery bays were not charging either battery.There was no patient involvement.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Manufacturer Narrative
Updated fields - b4,d9,g3,g6,h2,h3,h4,h6(type of investigation,investigation findings,component code,investigation conclusions,h10,h11.Corrected fields - e1(initial reporter,event site email),e2,e3.A getinge field service engineer (fse) had evaluated the unit and it was being found that neither battery bays charging either battery.Than the problem had been troubleshooted further from which the fse had replaced the pcba, cardiosave rohs, power management board from which both the battery bays are being charging.After the replacement the product had passed all the functional and safety tests per factory specifications.The defective components were received for further investigation.The failure analysis and testing dept.Received part number 0670-00-1162 rev.B, serial number: (b)(6) swemco with a reported unit failure of the batteries not charging.The fat performed a visual inspection and found the part to have a damaged q27 component.Fat will not install and test this part due to the risk to the test fixture.Fat verified that the board was faulty.This revision is obsolete and not able to be tested by the supplier.Retaining the part in the failure analysis and testing department per procedure number 0002-07-d008 rev.Aq.The non-conformances with the returned components were confirmed.However, the root cause or the most probable root cause is impossible to be defined.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key18546115
MDR Text Key333277679
Report Number2249723-2024-00280
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received04/30/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/28/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-