• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP SCISSORS; SCISSORS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP SCISSORS; SCISSORS, OPHTHALMIC Back to Search Results
Catalog Number 706.52
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/26/2023
Event Type  malfunction  
Event Description
A nurse reported that during a cataract surgery an ophthalmic scissors broke inside the eye and removed on the same day.The surgery was completed with no report of patient harm.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The sample received with inner blister, outer blister and cover foil showing the lot information.The sample shows macroscopic signs of damage.One blade of the scissors is broken.The sample shows no signs of surgery residues.The sample was visually inspected with the aid of a photomicroscope with various magnifications.The sample was visually inspected.It was noticed that one blade is broken.As the blister was opened by the customer, he handled the product.The point in time when the malfunction occurred, can no longer be determined.The customer¿s complaint is confirmed.It cannot be determined how or where the damage to the sample occurred; therefore, a root cause for the customers reported event could not be determined.The exact root cause for the customer¿s reported event is unknown, therefore, specific action cannot be taken.Complaints are reviewed and monitored at regular intervals for adverse trends.No adverse trends have been observed associated with the reported product and event.No action has been identified for this reported event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER REVOLUTION DSP SCISSORS
Type of Device
SCISSORS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18550841
MDR Text Key333320950
Report Number3003398873-2024-00011
Device Sequence Number1
Product Code HNF
UDI-Device Identifier07612717066885
UDI-Public07612717066885
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number706.52
Device Lot Number15FME0
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-