• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE; DEVICE IMPOTENCE MECHANICAL/HYDRAULIC Back to Search Results
Model Number 72404238
Device Problem Migration (4003)
Patient Problems Erosion (1750); Pain (1994)
Event Date 01/06/2024
Event Type  Injury  
Manufacturer Narrative
There was no device available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported patient symptoms are a known risk associated with implant of these devices as indicated in the instructions for use (ifu).Based on the information available, a conclusion code of known inherent risk of device was assigned to this investigation.
 
Event Description
It was reported that the cylinder of this inflatable penile prosthesis extruded through the meatus, causing pain for the patient.It is believed the extrusion was related to light modeling to reduce penile curvature due to peyronies disease performed post-implant by the physician.A revision surgery was performed to explant and plugged the cylinder, and another implant will be performed once the healing process is complete.No additional patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18553464
MDR Text Key333374265
Report Number2124215-2024-02978
Device Sequence Number1
Product Code FHW
UDI-Device Identifier00878953003511
UDI-Public00878953003511
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
N970012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number72404238
Device Catalogue Number72404238
Device Lot Number1100356616
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/06/2024
Initial Date FDA Received01/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/11/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
RESERVOIR: UPN: 720185-01 / LOT: 1100366202
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age61 YR
-
-