It was reported that the health care professional (hcp) called technical services (ts) and requested review of this cardiac resynchronization therapy defibrillator (crt-d) stored ventricular episodes.It was noted the patient does not have an implanted right atrial (ra) lead and may have received inappropriate atp therapy due to atrial fibrillation (af) with rapid ventricular response (rvr).Upon further review of the stored electrogram (egm), it was observed in the ventricular events, one of the episodes showed inappropriate atp therapy delivery appeared to have accelerated and induced the ventricular rhythm from ventricular tachycardia (vt) to a true ventricular fibrillation (vf) which was then converted with shock therapy.In the other ventricular episode, the vt rate zone was accelerated into the vf zone which led to inappropriate shock therapy to be delivered.The presenting egm was observed to have noisy signals in the left ventricular (lv) channel.Ts recommended for further device evaluation.The device remains in service.No additional adverse patient effects were reported.
|