• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO GIA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number GIA8038S
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2024
Event Type  malfunction  
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during an open procedure, there was difficulty or an inability to load.The two halves would not join and lock in place.A new device was used to resolve the issue.There was no patient injury.
 
Manufacturer Narrative
Additional information: d2, d9, g3, g4 (510k), h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the single use loading unit (sulu) was reset and loaded into the instrument.The sulu unloaded and loaded repeatedly without difficulty.The sulu and instrument were applied to test media with acceptable results.The firing knob of the instrument retracted and advanced properly without difficulty.The knife cut completely and cleanly and the staple line was properly formed.Hinge pin aligned with acceptable results.It was reported that there was difficulty with hinge pin alignment.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur due to improper loading of the cartridge.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: for properly loading the sulu into the instrument, which includes holding the sulu by the proximal end tabs and inserting into the cartridge fork at a 30 to 45 degree angle from the distal end down until the unit snaps in place.The shipping wedge is removed after the sulu is fully loaded.The sulu will "float" up and down in its final position.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GIA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key18554996
MDR Text Key333391676
Report Number2647580-2024-00441
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884523007268
UDI-Public10884523007268
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K221013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/29/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGIA8038S
Device Catalogue NumberGIA8038S
Device Lot NumberP3D0080
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/05/2024
Initial Date FDA Received01/22/2024
Supplement Dates Manufacturer Received02/22/2024
Supplement Dates FDA Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-