BOSTON SCIENTIFIC CORPORATION SYNERGY; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 10619
Device Problems Break (1069); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/23/2023
Event Type  malfunction  
Event Description
It was reported that shaft break occurred.The 85% stenosed target lesion was located in the moderately tortuous and moderately calcified left anterior descending artery.A 3.00 x 20 synergy drug-eluting stent was advanced for treatment.However, while crossing the tortuous anatomy and classified lesion, the shaft had been broken 15 to 20 cm from the hub towards the proximal end of the device.The device was simply pulled out from the patient intact.A 3 x 28mm synergy drug-eluting stent was then deployed, and the procedure was completed successfully.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18555146
MDR Text Key333393079
Report Number2124215-2024-02344
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10619
Device Catalogue Number10619
Device Lot Number0031602419
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age84 YR
Patient SexMale
Patient RaceAsian
-
-