Catalog Number CV-17702-E
Device Problem Obstruction of Flow (2423)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported "the catheter was found block during use on the patient".No medical intervention required.Pt condition reported as "fine".
 
Event Description
It was reported "the catheter was found block during use on the patient".No medical intervention required.Pt condition reported as "fine".
 
Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.The ifu provided with the kit informs the user, "flush each lumen with sterile normal saline for injection to establish patency and prime lumen(s)".A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CVC SET: 2-LUMEN 7 FR X 20 CM
Type of Device
CATHETER PERCUTANEOUS
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL CR, A.S.
jamska 2359/47
zdar nad sazavou 591 0 1
EZ   591 01
Manufacturer Contact
kevin don bosco
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key18557828
MDR Text Key333451294
Report Number3006425876-2024-00057
Device Sequence Number1
Product Code DQY
UDI-Device Identifier00801902105455
UDI-Public00801902105455
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K862056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberCV-17702-E
Device Lot Number71F21L1861
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/26/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NONE REPORTED.; NONE REPORTED.
-
-