B5 executive summary - updated.F10 / h6 health impact code grid - health effect - impact code ¿ updated.H4 manufacturing date ¿ added.H3 device evaluated by mfg ¿updated.H3 summary attached - updated.Section d catalog#, product long description, and gtin# - corrected.D4 expiration date - added.D9 product available to stryker ¿ updated.D9 returned to manufacturer on ¿updated.Due to the automated (manufacturing execution system) mes system there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection the coil delivery wire was seen to be kinked/bent in multiple areas.The main coil was seen to be stretched.The main coil suture was seen to be damaged.The coil delivery wire was seen to be broken/fractured.The coil introducer sheath was seen to be damaged.A mandrel was advanced through the returned catheter.Using a blade the catheter was then cut open and most of the main coil was retrieved from the catheter.Functional testing was not performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported event was not confirmed during analysis; however, the device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.An assignable cause of handling damage will be assigned to the analyzed damage of the coil delivery wire broken/fractured during use and to coil delivery wire kinked/bent, as the issue is due to handling of the product or portion of the product during the clinical procedure, upon removal of the product from the packaging, or preparation of the product prior to use.An assignable cause of procedural factors was assigned to the analyzed coil introducer sheath damaged as well as to the main coil suture damaged and main coil stretched as this complaint appears to be associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu, but performance was limited due to procedural factors during use.An assignable cause of not confirmed will be assigned to the reported event of the main coil prematurely detached/separated during use and coil jammed as the event was not confirmed during the analysis.
|