• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) UNIFY ASSURA CRT-D RF HV; NO MATCH Back to Search Results
Model Number CD3357-40C
Device Problem Insufficient Information (3190)
Patient Problem Syncope/Fainting (4411)
Event Type  Injury  
Manufacturer Narrative
Further information was requested but not received.
 
Event Description
Related manufacturer reference number: 2017865-2024-01499.It was reported the patient presented with syncope and an implantable cardioverter defibrillator (icd) that exhibited an unspecified malfunction.The left ventricular lead exhibited noise.The implantable cardioverter defibrillator (icd) was explanted and replaced, and the left ventricular lead was capped and replaced on (b)(6) 2023.The patient condition was unknown.
 
Manufacturer Narrative
The reported field event of sensing anomaly could not be replicated in the lab.Telemetry, impedance, sensing, pacing, and high voltage (hv) output functions of the device were tested and found to be normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY ASSURA CRT-D RF HV
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18560052
MDR Text Key333433586
Report Number2017865-2024-01498
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734508186
UDI-Public05414734508186
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Model NumberCD3357-40C
Device Lot NumberA000034603
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/22/2024
Supplement Dates Manufacturer Received01/26/2024
Supplement Dates FDA Received02/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age86 YR
Patient SexFemale
Patient Weight60 KG
-
-