ST. JUDE MEDICAL, INC.(CRM-SYLMAR) TENDRIL STS; NO MATCH Back to Search Results
Model Number 2088TC/52
Device Problem Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2023
Event Type  Injury  
Event Description
It was reported that the patient presented for an explant procedure.The right ventricular (rv) lead was explanted and replaced due to unknown reasons.The patient condition was unknown.
 
Manufacturer Narrative
The lead was returned without a reported event.As received, only the connector portion of the lead was returned for analysis.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspections of the partial lead did not find any anomalies except for procedural damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18560807
MDR Text Key333433548
Report Number2017865-2024-01642
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502887
UDI-Public05414734502887
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2022
Device Model Number2088TC/52
Device Lot NumberA000085291
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/03/2024
Initial Date FDA Received01/23/2024
Supplement Dates Manufacturer Received02/05/2024
Supplement Dates FDA Received02/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TENDRIL STS LEAD.
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-