• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUARTET; NO MATCH Back to Search Results
Model Number 1458Q/86
Device Problems Break (1069); Separation Failure (2547)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2024
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the stylet had failed to be separated from the left ventricular (lv) lead and an attempt to separate the stylet from the lead had resulted in the inner cable of the lv lead to come out.The lead was not used, and a new lead was implanted.The patient was stable throughout the procedure.
 
Manufacturer Narrative
The reported events of stylet could not be removed and ¿inner cable came back with the stylet¿ were confirmed.As received, a complete lead was returned with the stylet stuck inside both portions.Visual inspection of the lead found that the connector pin and crimp sleeve were pulled out of the connector assembly along with the inner coil which is consistent with procedural damage.Additionally, the ptfe stylet coating was bunched up/clogged with the inner coil distal to the connector pin which prevented the removal of the stylet.Excessive forces had resulted in the connector pin and crimp sleeve to be pulled out of the connector assembly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18561811
MDR Text Key333445016
Report Number2017865-2024-01649
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1458Q/86
Device Lot NumberA000149095
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/04/2024
Initial Date FDA Received01/23/2024
Supplement Dates Manufacturer Received02/05/2024
Supplement Dates FDA Received02/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUADRA ALLURE; RIGHT ATRIAL LEAD; RIGHT VENTRICULAR LEAD
Patient Age73 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-