• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR Back to Search Results
Model Number 00-3020
Device Problem No Pacing (3268)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2021
Event Type  malfunction  
Event Description
This report is based on information provided by philips remote service engineer (rse) and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus ls manual indicating that the device failed to pace during a training exercise with a "dpm hardware failure" message appearing.Remote service was provided and the customer requested a replacement device.A replacement device was sent out.The complaint was escalated for technical investigation.Although the initial entry tests were passed and no malfunction was detected, the log files were extracted and investigated.Error 26 appeared in the log file.The root cause could not be determined as the error was not reproducible.Based on the information available and the testing conducted, the cause of the reported problem cannot be determined however the likely cause is the board-to-board connector between the dpm board and the mainboard.The reported problem was confirmed.Based on the information available and results of additional analysis further action has been initiated.An analysis was performed by a vigilance reporting specialist (vrs).The vrs determined that while no harm was alleged, recurrence of this failure mode during clinical use could cause or contribute to death or serious injury.Therefore, this failure mode meets the definition of a reportable malfunction.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The device was replaced to resolve the issue.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Manufacturer Narrative
Updated conclusion grid and component grid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS LS-MANUAL
Type of Device
LOW ENERGY DEFIBRILLATOR
Manufacturer (Section D)
SCHILLER AG
altgasse 68
baar 6341
SZ  6341
Manufacturer (Section G)
SCHILLER AG
altgasse 68
baar 6341
SZ   6341
Manufacturer Contact
tanya deschmidt
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key18564913
MDR Text Key333500023
Report Number3003832357-2024-00048
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-3020
Device Catalogue Number989706001681
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/28/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-