• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO PROCUITY; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO PROCUITY; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Catalog Number 300900000000
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/01/2023
Event Type  malfunction  
Event Description
This report summarizes 5 malfunction events, where it was reported the devices experienced accessible ac current.There was no patient involvement.
 
Manufacturer Narrative
This record is a consolidation of records summarized as part of the fda voluntary malfunction summary reporting program.5 devices were functionally/visually inspected in the field.The devices were repaired and returned to use.There was no remedial action taken.This device is not labeled for single use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCUITY
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
melissa simon
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key18565629
MDR Text Key333505872
Report Number0001831750-2024-00169
Device Sequence Number1
Product Code FNL
UDI-Device Identifier07613327512656
UDI-Public07613327512656
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported5
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number300900000000
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 01/01/2024
Initial Date FDA Received01/23/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-