ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY ASSURA DR ICD, US; Pulse generator, permanent, implantable Back to Search Results
Model Number CD2357-40Q
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem Discomfort (2330)
Event Type  malfunction  
Event Description
It was reported that the implantable cardioverter defibrillator was found in backup vvi mode due to event queue overflow.The patient was mildly symptomatic.The device was restored successfully.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18567073
MDR Text Key333519947
Report Number2017865-2024-01708
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734508094
UDI-Public05414734508094
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 01/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2018
Device Model NumberCD2357-40Q
Device Lot NumberA000030790
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/06/2024
Initial Date FDA Received01/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/23/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2088TC/52 TENDRILLEADS; 7122Q/58 DURATAACTIVE LEAD
Patient Age71 YR
Patient SexFemale
-
-