• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD VENTED AUTOFEED HUMIDIFICATION CHAMBER; BTT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD VENTED AUTOFEED HUMIDIFICATION CHAMBER; BTT Back to Search Results
Model Number MR290
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/26/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The subject mr290v vented autofeed humidification chamber has been requested to be returned to fisher & paykel (f&p) healthcare for evaluation.We will provide a follow up report upon completion of investigation.
 
Event Description
A distributor in (b)(4) reported that a mr290v vented autofeed humidification chamber was found cracked before patient use.There was no reported patient involvement.
 
Event Description
A distributor in (b)(6) reported that a mr290v vented autofeed humidification chamber was found cracked before patient use.There was no reported patient involvement.
 
Manufacturer Narrative
(b)(4).Corrected data: section d2 - common device name: corrected to btt.Method: the subject mr290v vented autofeed humidification chamber was not returned to fisher & paykel healthcare for evaluation.Our investigation is thus based on the information and photographs provided by the customer, and our knowledge of the product.Results: visual inspection of the provided photograph confirmed a vertical crack in the wall of the chamber dome.Conclusion: the complaint device was requested for investigation, however it was not provided for analysis.Without the return of the complaint device, we are unable to determine the cause of the reported event.Every mr290v chamber is pressure tested following the manufacturing process to check for any leaks present in the chamber due to cracks and other causes.Any chamber which fails this test is rejected.In addition, the pressure test is followed by a visual inspection of each chamber.No cracks in the chamber dome are acceptable.Any chamber that fails this inspection is rejected.The subject mr290v chamber would have met the required specification at the time of production.Our user instructions that accompany the mr290v vented autofeed humidification chamber state the following: "do not soak, wash, sterilize, or reuse this product.Avoid contact with chemicals, cleaning agents, or hand sanitizers." "ensure appropriate ventilator or flow source alarms are set before connecting breathing set to patient." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "use of the mr290 above the maximum operating pressure may lead to cracking, water leakage and, on rare occasions could lead to a loss of ventilation pressure.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENTED AUTOFEED HUMIDIFICATION CHAMBER
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
omid taheri
17400 laguna canyon road
suite 300
irvine, CA 92618
8007923912
MDR Report Key18567936
MDR Text Key333540591
Report Number9611451-2024-00044
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K934140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/28/2023
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR290
Device Catalogue NumberMR290
Device Lot Number2102228803
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/28/2023
Initial Date FDA Received01/23/2024
Supplement Dates Manufacturer Received02/15/2024
Supplement Dates FDA Received03/07/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/20/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-