Model Number OER-AW |
Device Problem
Device Reprocessing Problem (1091)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/31/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus, the circulation port filters of the endoscope reprocessor were damaged.No error code was displayed.The issue occurred during reprocessing for a tracheoscopy procedure.There was no delay, and the patient was not under anesthesia.The intended procedure was completed with a similar device.The device was returned for evaluation and the following reportable malfunction was found: as the circulation inlet mesh filter is used to filter foreign matter, in the event of damage, foreign matter could clog the pipeline and cause reprocessing failure.There were no reports of patient harm or impact associated with the event.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
|
Manufacturer Narrative
|
The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause of the reported circulation port mesh filter issue could not be determined, however, it is possible that part of the scope hit the filter when settingup the scope or an accessories such as a cleaning tube was dropped onto the filter, causing damage.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|