Model Number 2088TC/58
Device Problems Difficult to Fold, Unfold or Collapse (1254); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2024
Event Type  malfunction  
Event Description
It was reported that the patient presented for an implant procedure.During the procedure, it was noted that the helix portion of the lead failed to extend, resulting in unacceptable parameters.The lead was not used and replaced.The patient was in stable condition.
 
Manufacturer Narrative
Further information was requested but not received.
 
Manufacturer Narrative
The reported event of helix mechanism issue was confirmed.A complete lead was returned in one piece.Electrical testing did not find any indication of conductor fractures.Short between conductors was found due to the over-torqued inner coil at the connector region.The lead was returned with the helix retracted and clogged with blood.X-ray inspection found the inner coil over-torqued at the connector region consistent with procedural damage.X-ray examination of the helix mechanism found no anomalies or distortion of the helix that would have contributed to helix mechanism issue reported in the field.The cause of the reported event was isolated to blood in the helix region and over-torqued of the inner coil.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
NO MATCH
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18568850
MDR Text Key333539349
Report Number2017865-2024-01768
Device Sequence Number1
Product Code NVN
UDI-Device Identifier05414734502894
UDI-Public05414734502894
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2088TC/58
Device Lot NumberP000183483
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/04/2024
Initial Date FDA Received01/24/2024
Supplement Dates Manufacturer Received02/22/2024
Supplement Dates FDA Received02/27/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-