Catalog Number 05.001.201 |
Device Problems
Break (1069); Appropriate Term/Code Not Available (3191); Noise, Audible (3273); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2024 |
Event Type
malfunction
|
Event Description
|
It was reported by india that during service and evaluation, it was determined that the trigger of the battery handpiece/modular device was sticky.It was further observed that the device would not run.It was further determined that the device failed pretest for check for mechanical free moving, check for sticky triggers and check general function of device.It was noted in the service order that the device was not working properly, the spring of the forward trigger was damaged and not completely depressing, and the device was noisy while running.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in (b)(6) 2024.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of the device having a component damage was not confirmed.Therefore, an assignable root cause was not determined.However, the sticky trigger, identified during service and evaluation was confirmed.The assignable root cause was determined to be due to component failure from normal wear.Udi: (b)(4).
|
|
Search Alerts/Recalls
|
|
|