• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRHYTHM TECHNOLOGIES, INC ZEUS SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRHYTHM TECHNOLOGIES, INC ZEUS SYSTEM; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Device Problem Failure to Transmit Record (1521)
Patient Problem Unspecified Heart Problem (4454)
Event Date 12/27/2023
Event Type  malfunction  
Manufacturer Narrative
The zio at device was returned to irhythm, and the clinical data was downloaded.A review of the clinical data found that the patient wore the zio at device for 11 days of the 14-day prescribed wear period.Irhythm became aware of the arrhythmia while preparing final report and notified the hcp on day 18.The investigation revealed a potential algorithm sensitivity issue with the device, as there were no errors or issues observed around the time the missed episode occurred.This event is being reported per 21cfr 803 as an product problem /malfunction.This report does not constitute an admission by irhythm that the product described in this report has any defects or has malfunctioned.These terms are included in form fda 3500a and are fixed terms for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.
 
Event Description
The patient experienced an arrhythmia that met medical doctor notification (mdn) requirements that was not transmitted during the wear period.The investigation revealed a potential algorithm sensitivity issue with the device.The healthcare provider (hcp) was immediately notified, and irhythm learned that the hcp was already aware of the patient's arrhythmia and was treating it.There were no delays in treatment, and no adverse events, such as death or serious injury, are known to have occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEUS SYSTEM
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
IRHYTHM TECHNOLOGIES, INC
699 8th st suite 600
san francisco CA 94103
Manufacturer (Section G)
IRHYTHM TECHNOLOGIES, INC
6550 katella avenue, suite 200
cypress CA 90630
Manufacturer Contact
mazi kiani
699 8th st suite 600
san francisco, CA 94103
9494132147
MDR Report Key18576904
MDR Text Key334499895
Report Number3007208829-2024-00034
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K222389
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2024
Date Manufacturer Received01/09/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ZIO AT/SN (B)(6).
Patient Age75 YR
Patient SexMale
-
-