Model Number 720185-01
Device Problems Inflation Problem (1310); Material Deformation (2976); Migration (4003); Air/Gas in Device (4062)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2023
Event Type  Injury  
Manufacturer Narrative
The device is not available for analysis; therefore, no physical or visual analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Based on the information available, the cause that contributed to the reported event cannot be established; a conclusion code of cause not established was assigned to this investigation.Model number: 72404252-10.Lot number: 1100369918.Catalog description: lgx preconnect ms 18cm ps 10cm tl iz.
 
Event Description
It was reported that this inflatable penile prosthesis (ipp) exhibited inflation issues due to reservoir migration from the retzius space to the patient's abdomen, no tissue perforation occurred due to the migration.In addition, the tubing was reported to be kinked and air was found inside the pump.A revision surgery was performed to replace only the reservoir and place a new connection.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS INFLATABLE PENILE PROSTHESIS WITH INHIBIZONE
Type of Device
DEVICE IMPOTENCE MECHANICAL/HYDRAULIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key18578769
MDR Text Key333667698
Report Number2124215-2024-03336
Device Sequence Number1
Product Code FHW
UDI-Device Identifier00878953005669
UDI-Public00878953005669
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
N970012
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number720185-01
Device Catalogue Number720185-01
Device Lot Number1100363393
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LGX18: UPN: 72404252-10 / LOT: 1100369918.
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age61 YR
-
-