ZIMMER SURGICAL, INC. UNKNOWN DERMATOME BLADE Back to Search Results
Device Problems Failure to Cut (2587); Patient Device Interaction Problem (4001)
Patient Problem Laceration(s) (1946)
Event Date 11/21/2023
Event Type  Injury  
Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.G2: canada.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Part and lot identification are necessary for review of device history records, neither were provided.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated report: 0001526350-2024-00143.
 
Event Description
It was reported that during surgery a deep wound (7cm long and 1cm deep) was created when pressing the dermatome on the left thigh.The harvest site required sutures and an additional skin graft was required to complete the surgery.There was an unspecified amount of surgical delay to repair the wound.Another dermatome was used to complete surgery.During device evaluation it was discovered that the blade may have contributed to the reported event.Due diligence is complete, there is no additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN DERMATOME BLADE
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key18579485
MDR Text Key333728161
Report Number0001526350-2024-00144
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 01/24/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2024
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Other;
Patient SexPrefer Not To Disclose
-
-