ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 14687-15 |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/25/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been received for evaluation.Investigation is pending.
|
|
Event Description
|
The event involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch where damaged tubing was reported.The event occurred during infusion.There was patient involvement but no patient harm.
|
|
Manufacturer Narrative
|
Received a photo from the customer showing tubing split in two.Received one used list #14687-15 primary plum set.As received, the tubing after the cassette and before the y-clave was observed to be cut.The complaint of tubing damage on the 14687-15 primary plum set can be confirmed.The manufacturing process was reviewed and there are no sharp objects or instruments on the manufacturing floor capable of the observed damage as a preventative measure.The complaint of tubing damage can be confirmed.The cause of the damage is unknown; however, it would have occurred outside of icu medical control.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
|
|