U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ICU MEDICAL COSTA RICA LTD. PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 14687-15
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/25/2023
Event Type  malfunction  
Manufacturer Narrative
The device has been received for evaluation.Investigation is pending.
 
Event Description
The event involved a primary plum set, clave secondary port, clave y-site, secure lock, 103 inch where damaged tubing was reported.The event occurred during infusion.There was patient involvement but no patient harm.
 
Manufacturer Narrative
Received a photo from the customer showing tubing split in two.Received one used list #14687-15 primary plum set.As received, the tubing after the cassette and before the y-clave was observed to be cut.The complaint of tubing damage on the 14687-15 primary plum set can be confirmed.The manufacturing process was reviewed and there are no sharp objects or instruments on the manufacturing floor capable of the observed damage as a preventative measure.The complaint of tubing damage can be confirmed.The cause of the damage is unknown; however, it would have occurred outside of icu medical control.The lot history was reviewed, no nonconformities were identified that may have contributed to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY PLUM SET, CLAVE SECONDARY PORT, CLAVE Y-SITE, SECURE LOCK, 103 INCH
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key18580220
MDR Text Key333822777
Report Number9615050-2024-00072
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10887787005650
UDI-Public(01)10887787005650(17)260501(10)13615680
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K141789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number14687-15
Device Lot Number13615680
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2024
Is the Reporter a Health Professional? No
Date Manufacturer Received02/07/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-