This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, there is a possibility that the event occurred due to physical stress caused by hitting the tip of the scope or dropping the tip, or chemical stress due to the chemical solution used, but it could not be identified.The event can be detected/prevented by following the instructions for use which state: "ltf-190-10-3d operation manual important information ¿ please read before use: contraindications, warnings, and precautions: do not strike, hit, or drop the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connectors, or light guide connector.Also, do not bend, pull, or twist the endoscope¿s distal end, rigid portion, bending section, control section, universal cord, video connectors, or light guide connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient.Ltf-190-10-3d reprocessing manual 3.1 compatibility summary: methods listed as ¿compatible¿ in table 3.1 and 3.2 are compatible for routine use only when used according to manufacturer¿s instructions.Repeated use and reprocessing of endoscopes and accessories lead to gradual wear and tear.Furthermore, reprocessing methods that employ higher temperatures and more caustic/corrosive materials may lead to faster deterioration.In general, sterilization processes are harsher on equipment than disinfection processes.Before each patient procedure, inspect the endoscope and accessories for damage, according to the instructions described in this manual and its companion ¿operation manual¿.Reprocessing manual: 3.1 compatibility summary: list of compatible methods validated in terms of material durability: sterrad® 50/100s/200/nx¿: sterilization by sterrad® 50/100s/200/nx¿ system may deteriorate the adhesive of the insertion section.Depending on the circumstances, replacement of the insertion section may be required.Before use, confirm that the endoscope is free from any damage or other irregularities.For further details, contact olympus".Olympus will continue to monitor field performance for this device.
|