• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION DYNAGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION DYNAGEN X4 CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D) Back to Search Results
Model Number G158
Device Problems Over-Sensing (1438); Telemetry Discrepancy (1629)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/23/2023
Event Type  malfunction  
Event Description
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) device recorded a signal artifact monitor (sam) event.There was difficulty interrogating device with the current software.Technical services (ts) stated to use the 3300 programmers to interrogate the device that have smr15 on them.Upon review, the presenting electrogram (egm) detected noise.Impedance measurements were within the range.The plant is to interrogate the device.This device remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN X4 CRT-D
Type of Device
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATORS (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18580746
MDR Text Key333737927
Report Number2124215-2024-03936
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534904
UDI-Public00802526534904
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/12/2021
Device Model NumberG158
Device Catalogue NumberG158
Device Lot Number225107
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/12/2024
Initial Date FDA Received01/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/25/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
Patient SexMale
-
-