• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II DUODENOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-Q180V
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/07/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for annual inspection, and the evaluation found the reportable malfunction of the distal end had a crack.The following non-reportable findings were found during device evaluation: grip had a scratch, air/water cylinder had no color, suction cylinder had no color, switch box had a scratch, right/left knob had discoloration, upward/downward angulation control knob had discoloration, water tightness was lost because guide pin of electrical connector was shaved, scope connector cover unit had a scratch, scope connector had a scratch, sheet holder unit had corrosion due to water leakage, light guide lens had a chip, adhesive around light guide lens had a crack, connecting tube had a scratch, adhesive around objective lens had a crack, and universal cord had a scratch.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
It was observed that during the device evaluation, the duodenovideoscope exhibited a crack on the distal end.There were no reports of patient involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was visually inspected, and functional testing was performed.The device did not meet the standard specifications and functions.Based on the results of the investigations, the gap between z-block and c-body was confirmed.The probable cause was determined as due to physical stress such as hitting/dropping distal end or chemical stress such as chemical solution used.However, the root cause of the suggested event could not be specified.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18583047
MDR Text Key333824072
Report Number9610595-2024-01640
Device Sequence Number1
Product Code FDT
UDI-Device Identifier04953170367311
UDI-Public04953170367311
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K143153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-Q180V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/25/2024
Supplement Dates Manufacturer Received02/15/2024
Supplement Dates FDA Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/07/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-