MAUDE Adverse Event Report: GYRUS ACMI, INC. GYRUS, PK-SP GENERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. GYRUS, PK-SP GENERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 744000
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/21/2023
Event Type  malfunction  
Event Description
It was observed that during the device evaluation, the gyrus, pk-sp generator exhibited a faulty electrode.There were no reports of patient involvement.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was not confirmed.The device evaluation found that there was a faulty electrode.The investigation is ongoing and followup with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and additional event information.The customer was contacted on multiple occasions for additional event information.No response was received.  since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.During investigation, the reported issue was not observed.The reported issue (faulty electrode) would likely be caused by the device failing the current limit test due to a faulty radiofrequency board.However, the issue could not be confirmed; the root cause is unknown.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GYRUS, PK-SP GENERATOR
Type of Device
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18584200
MDR Text Key334798671
Report Number3011050570-2024-00024
Device Sequence Number1
Product Code GEI
UDI-Device Identifier00821925009271
UDI-Public00821925009271
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number744000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2024
Was the Report Sent to FDA? No
Date Manufacturer Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-