GYRUS ACMI, INC. GYRUS, PK-SP GENERATOR; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
Model Number 744000 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/21/2023 |
Event Type
malfunction
|
Event Description
|
It was observed that during the device evaluation, the gyrus, pk-sp generator exhibited a faulty electrode.There were no reports of patient involvement.
|
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation and the customer's allegation was not confirmed.The device evaluation found that there was a faulty electrode.The investigation is ongoing and followup with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and additional event information.The customer was contacted on multiple occasions for additional event information.No response was received. since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.During investigation, the reported issue was not observed.The reported issue (faulty electrode) would likely be caused by the device failing the current limit test due to a faulty radiofrequency board.However, the issue could not be confirmed; the root cause is unknown.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|