• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE 2; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71953-01
Device Problem Display or Visual Feedback Problem (1184)
Patient Problems Hypoglycemia (1912); Loss of consciousness (2418)
Event Date 12/29/2023
Event Type  Injury  
Event Description
A customer reported a display issue with the adc device.The customer observed "frozen" display and was unable to view glucose readings.The customer became hypoglycemic and lost consciousness, requiring treatment of glucagon injection administered by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
 
Manufacturer Narrative
The reported device has been returned and is currently in process of investigation.A follow-up report will be filed once additional information is obtained.The date of event is unknown.The date entered in section b3 is the date abbott diabetes care became aware of the event.The device manufacturing date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.The customer indicated this event occurred three times, this report covers 2 of 3 reports.All pertinent information available to abbott diabetes care has been submitted.
 
Manufacturer Narrative
The reader (b)(6) was returned and investigated.Visual inspection has been performed and liquid contamination was observed in the usb port.The liquid contamination in the usb port would prevent the customer from charging the reader which would lead to the reader not turning on.The reader was de-cased.Visual inspection was performed and liquid contamination was observed due to liquid ingress on the pcba(printed circuit board assembly).Therefore, issue in not confirmed to use due to liquid contamination.An extended investigation has been performed for the reported complaint.There was no indication that the product did not meet specification.The dhrs (device history review) for the libre reader were reviewed and the dhrs showed the libre reader passed all tests prior to release.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
A customer reported a display issue with the adc device.The customer observed "frozen" display and was unable to view glucose readings.The customer became hypoglycemic and lost consciousness, requiring treatment of glucagon injection administered by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE 2
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
Manufacturer Contact
audra fuentes
1360 south loop road
alameda, CA 94502-7001
5107495297
MDR Report Key18584535
MDR Text Key333771720
Report Number2954323-2024-03314
Device Sequence Number1
Product Code QLG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K193371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71953-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2024
Initial Date Manufacturer Received 12/29/2023
Initial Date FDA Received01/25/2024
Supplement Dates Manufacturer Received02/16/2024
Supplement Dates FDA Received02/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/02/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-